Important Points to Note

  • Store the test kit at 2°C–30°C and do not freeze it.
  • Before testing, allow the test device, specimen and buffer to reach room temperature (15°C–30°C).
  • Use the test device as soon as possible after opening the foil pouch.
  • For a finger-prick whole blood sample, add 2 drops of blood to the sample well, followed by 1 drop of test buffer.
  • Read the result at 15 minutes. Do not interpret the result after 15 minutes.

Reading Your Results

Important

Read the result at 15 minutes. Do not interpret the result after 15 minutes. Always follow the Instructions for Use supplied with the test.

Reactive result

Two coloured lines are visible in the testing cartridge: one in the control (C) area and one in the test (T) area. The test line may vary in colour intensity. Any visible test line together with a valid control line should be interpreted as a reactive result.

A reactive HIV self-test result does not by itself confirm an HIV diagnosis. Appropriate confirmatory testing by a healthcare professional or testing service is required.

Non-reactive result

One coloured line is visible in the control (C) area and no coloured line is visible in the test (T) area. This is interpreted as a non-reactive result.

A non-reactive result does not exclude HIV infection if testing was performed during the applicable window period following a recent possible exposure. Depending on when the possible exposure occurred, repeat testing may be necessary.

Invalid result

The result is invalid if no control (C) line appears, regardless of whether a line is visible in the test (T) area. Do not interpret an invalid result.

Repeat the test using a new test device and follow the Instructions for Use carefully. If the problem persists, discontinue use and follow the manufacturer's guidance or contact the distributor.